        Type Selection Instructions for Micro Flow Metal Tube Rotor Flowmeter
        • File Name:Type Selection Instructions for Micro Flow Metal Tube Rotor Flowmeter
        • Add Time:2019-04-12 10:30:20
        • Document Size:
        • Recommendation:
        • Immediately Download:
        Download Instructions:

        Productoverview:
        ZW-LZZW series of micro flow of metal tube rotameter has the characteristics of strong, stable, reliable and wide application range, can be used for measuring gas and liquid.All metal design to ensure its high pressure, high temperature stability, can withstand extreme environmental conditions.The position of the magnetic induction transmitted to an analog indicator displays the measured values.Without power.
        Two, the structure and principle of measurement:
        1, structure:
        ZW-LZZW series of micro flow of metal tube rotameter is mainly composed of three parts:
        A (Intelligent indicator, indicator, local indicator);
        B, float;
        C, a conical measuring chamber.
        2, the principle of measurement:
        On the lower end of the measured medium from the measuring tube through a conical float generating differential pressure form the upward force, when the float on the lift is greater than immersion float in the fluid weight, the float will rise, annulus area increases, the annulus fluid velocity decreased immediately, float on the lower pressure reduced, to float up with the decrease in stress, until the lift is equal to the weight of the float when immersed in a fluid, the float will stabilize at a certain height.The float position that corresponds to the size of the measured medium flow.A built-in magnet, the float with the medium moves up and down, with the movement of the float magnetic field changes.
        A, for in situ, by rotary magnetic float and the local indicator in the inner magnetic coupling, rotate, and drives the pointer, through the dial indicates the size of the flow.
        B, the intelligent, by a solid state magnetic sensor intelligent indicator in the magnetic field will change into electrical signal output by A/D transform, D/A microprocessor, LCD liquid crystal display to show the size and cumulative flow flow.
        Three, widely used:
        ZW-LZZW series of micro flow of metal tube rotameter is widely used in petroleum, chemical, power, pharmaceutical, food, water treatment, etc..The complex, harsh environmental conditions, the flow measurement process and various medium conditions in.
        Fourth, the main features:
        1, all metal structure, strong stability, suitable for high temperature and high pressure, strong corrosive medium.
        2, using the isolation design, imported relay output, to ensure reliable operation.
        3, in all sectors of the measurement of gas and liquid, the measuring part can be using different materials to adapt to various medium conditions.
        4, the vertical and horizontal installation, more suitable for different occasions.
        5, the magnetic coupling structure to ensure data transmission, more stable signal.
        6, magnetic filter, with thermal insulation or jacket design.
        7, the vertical and horizontal installation, more suitable for different occasions.
        8, can be used in flammable and explosive occasions.
        Five, technical parameters:
        1. measuring range: water (20 C) (0.3 ~ 100) L/h; air (20 degrees 101325Pa) (1.2 ~ 3400) L/h;
        2. range: 10:1;
        3. accuracy level: 4
        Maximum working pressure: 4. 2.5MPa (according to the need of increase)
        5. medium temperature: -80 ~ +150 C
        6. environment temperature: -25 ~ +65 C
        7. connection: 1/4 "NPT thread or phi 6 (phi 8) sleeve connection / vertical connection according to the level of user requirements
        8. limit alarm: 1~2 alarm switch (optional)
        9. dimensions: overall size: 110mm; installation center distance: 90mm;
        Sixth, product maintenance:
        1, in the process of long-term use, the pipeline inevitably there will be ferromagnetic impurity adsorption on the float, it will float stuck or affect the measurement accuracy, from time to time to clean the sensor float; if the installation of magnetic filters should be cleaned regularly.
        2, instrument indicator equipped with electronic components, instrument wiring or demolition of housing, to housing screws, must ensure sealing, prevent impurities, or other material into the water, and also to ensure reliable earthing of the meter casing.
        3, instrument installation, on first use, should pay attention to:
        When the valve is opened to avoid the formation of pressure shock makes the float slamming limiting device, causing damage to the instrument, slowly open the valve!
        The measurement of gas meter is equipped with gas damper, to minimize the shock; to ensure the stability of the float, a throttle valve can be installed in the instrument export.
        4, for the remote type instrument, we must first ensure that the instrument wiring correctly, keeps power; for dangerous situations, must choose the explosion-proof, and the installation and use of proof required by the.
        Seven, the installation should follow the following steps:
        1, before installing the instrument to process piping, should remove all installed and checked for any transport damage;
        2, in order to prevent in transit, with a rubber band fixed pointer indicator, indicator should open the left cover it gently removed before connection, do not change the pointer position;
        3, open the indicator of the right cover, cable through the cable introducing device, according to the wiring diagram wiring rotation transmitter, pressure screw nut, the cable is fixed, and then tighten the fastening screws around the cover, the indicator of waterproof and dustproof;
        4, upstream and downstream piping and instrument instrument should be the same caliber, flange or threaded connection should be matched with the flange and screw instrument, instrument upstream straight pipe length should be 5 times the nominal diameter of the instrument, the downstream straight pipe length not less than H2 250mm;
        5, the instrument measuring mechanism is magnetic transfer, so in order to ensure the performance of rotameter installed around the sensor, at least 100mm, the existence of ferromagnetic material is not allowed;
        Before installation, 6 meter pipeline sweeping shall be carried out, to prevent the retention of ferromagnetic material in the pipeline is attached to the instrument, the instrument performance, or even damage the instrument. If unavoidable, should be installed at the entrance of the magnetic filter instrument;
        7, the measurement of gas meter is calibrated in special pressure, if the gas meter in the outlet directly discharged into the atmosphere, will produce a pressure drop in the float, and lead to data distortion.If this is the working conditions, should install a valve in the instrument exports, so that the required flow value is set, when the float to maintain calibration pressure, gas expansion in valve;
        The installation form 8, instrument installation for vertical and horizontal installation, if it is installed in a vertical straight form, should ensure that the instrument is better than 1%; if the level of installation form should ensure that the level of less than 1% meters;
        9, installed in the pipeline in the instrument should not be subject to stresses, the instrument should be a proper entrance and exit pipe support, can make the instrument in the minimum stress state;
        10, product and computer related to calibration and other operations, must be in a safe place.
        Eight, installation notes:
        In 1, before the first installation flowmeter in the pipeline, dirt, impurities such as welding blowing clean, if the liquid contained solid impurities, should the corresponding filter is installed between the valve and the straight pipe.
        2, shielding and grounding system: in order to ensure ideal electromagnetic compatibility, shielding and grounding is very important, especially the bus cable, cable as far as possible to make continuous shielding.
        3, the safety zone, according to the requirements of conventional cable grounding as much as possible.
        4, in dangerous places, to ensure that the whole range of bus safe places and dangerous enough, all the potential, in this case, multiple ground is very superior.

        Nine, choose the flow table:

        Decay path

        Common path

        measuring range

        The maximum pressure loss

        DN (mm)

        DN (mm)

        The air (m3/h)

        Water (L/h)

        The air (KPa)

        Water (KPa)

        Fifteen

        Fifteen

        0.05-0.5

        1.6-16

        Seven

        Six point four

        0.07-0.7

        2.5-25

        Seven point one

        Six point five

        0.11-1.1

        4.0-40

        Seven point two

        Six point five

        0.18-1.8

        6.0-60

        Seven point three

        Six point six

        0.28-2.8

        10-100

        Seven point five

        Six point six

        0.40-4.0

        16-160

        Eight

        Six point eight

        0.70-7.0

        25-250

        Ten point eight

        Seven point two

        1.00-10

        40-400

        Ten

        Eight point six

        1.60-16

        60-600

        Fourteen

        Eleven point one

        Twenty-five


        80-800


        Fourteen

        Twenty-five

        3.00-30

        100-1000

        Seven point seven

        Seven

        4.50-45

        160-1600

        Eight point eight

        Eight

        7.00-70

        250-2500

        Twelve

        Ten point eight

        11-110

        400-4000

        Nineteen

        Fifteen point eight

        Fifty


        500-5000


        Sixteen point two

        Fifty

        18-180

        600-6000

        Eight point six

        Eight point one

        25-250

        1000-10000

        Ten point four

        Eleven

        Eighty

        40-400

        1600-16000

        Twelve point six

        Seventeen

        Eighty

        50-500

        2000-20000

        Fifteen point five

        Six point three

        One hundred

        75-750

        2500-25000

        Seventeen Point Two

        Eight point one

        One hundred

        100-1000

        4000-40000


        Nine point five

        One hundred and twenty-five

        160-1600

        6000-60000


        Ten

        One hundred and twenty-five

        250-2500

        8000-80000


        Sixteen

        One hundred and fifty


        10000-100000


        Twenty


        300-3000

        12-120m3/h


        Twenty


        400-4000

        15-150m3/h


        Twenty-five

        Ten, selection table:

        Specification code

        Explain

        The name of the factory

        ZW

        Zhongwei control

        Instrument types

        LZZW

        The micro flow of metal tube flowmeter

        structural style

        A

        In the next

        B

        Progress under the

        C

        Side measure

        D

        The bottom side

        R

        Right into the left out

        L

        The left into the right

        K

        Clamp connection (sanitary polishing tube)

        W

        Threaded connection

        Size code

        - XXX

        DN15 said: 15 cases

        The measuring tube material

        R1

        1Cr18Ni9Ti, 304

        R0

        0cr18Ni12M02Ti, 316

        RL

        316L

        RP

        PTFE (PTFE)

        Ni

        Hastelloy (Hastelloy C)

        Ti

        TitHXium (Tai Hejin)

        Media type

        L

        liquid

        G

        Gas

        working temperature

        H1

        - 40 -100 OC

        H2

        - 80 -250 OC

        Rated pressure

        One point six

        1.6MPa

        Four

        4.0MPa

        Sixteen

        16MPa

        Twenty-five

        25MPa

        Additional structure

        O

        nothing

        Z

        Damping type

        T

        Jacket

        H

        High temperature type

        Y

        High pressure type

        Form indicator

        M1

        Local indication

        M2

        Intelligent

        M3

        Intelligent (explosion)

        The output signal in the communication

        N

        nothing

        A

        4-20mA

        H

        4-20mA+Hart

        Alarm controller

        K1

        The upper limit alarm

        K2

        Alarm

        K12

        The upper and lower limit alarm

        Explosion proof type

        E1

        Intrinsically safe EXiaCT5 square casing

        E2

        Flameproof shell EXiibt4 Park

        Power supply

        Zero

        24VDC (optional)

        Two

        220VAC

        Three

        The 3.6 volt lithium battery